 
| 48T | ¥1960.00 | 
| 96T | ¥2800.00 | 
| 96T*5 | ¥11900.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 100.00 | 2.241 | 2.145 | 
| 50.00 | 1.635 | 1.539 | 
| 25.00 | 1.222 | 1.126 | 
| 12.50 | 0.906 | 0.810 | 
| 6.25 | 0.636 | 0.540 | 
| 3.13 | 0.385 | 0.289 | 
| 1.56 | 0.210 | 0.114 | 
| 0.00 | 0.096 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 93-119% | 106% | 
| EDTA plasma(n=5) | 94-99% | 97% | 
| Heparin plasma(n=5) | 84-103% | 94% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-117% | 80-88% | 82-91% | 91-97% | 
| EDTA plasma(n=5) | 94-107% | 92-110% | 88-103% | 88-100% | 
| Heparin plasma(n=5) | 92-113% | 89-97% | 94-103% | 85-117% | 

 
            关注微信公众号