| 48T | ¥1960.00 | 
| 96T | ¥2800.00 | 
| 96T*5 | ¥11900.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 1.952 | 1.859 | 
| 1000.00 | 1.622 | 1.529 | 
| 500.00 | 1.099 | 1.006 | 
| 250.00 | 0.880 | 0.787 | 
| 125.00 | 0.579 | 0.486 | 
| 62.50 | 0.325 | 0.232 | 
| 31.25 | 0.224 | 0.131 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 89-102% | 96% | 
| EDTA plasma(n=5) | 88-111% | 100% | 
| Heparin plasma(n=5) | 93-110% | 102% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 93-99% | 89-112% | 81-87% | 92-106% | 
| EDTA plasma(n=5) | 90-99% | 87-95% | 84-90% | 84-99% | 
| Heparin plasma(n=5) | 87-105% | 82-92% | 86-98% | 91-117% | 

 
            关注微信公众号