| 48T | ¥1960.00 |
| 96T | ¥2800.00 |
| 96T*5 | ¥11900.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.978 | 1.880 |
| 5.00 | 1.643 | 1.545 |
| 2.50 | 1.259 | 1.161 |
| 1.25 | 0.799 | 0.701 |
| 0.63 | 0.515 | 0.417 |
| 0.31 | 0.389 | 0.291 |
| 0.16 | 0.228 | 0.130 |
| 0.00 | 0.098 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-91% | 87% |
| EDTA plasma(n=5) | 80-129% | 105% |
| Heparin plasma(n=5) | 90-103% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-102% | 88-120% | 81-98% | 89-105% |
| EDTA plasma(n=5) | 82-112% | 83-116% | 92-106% | 90-99% |
| Heparin plasma(n=5) | 94-113% | 88-117% | 92-99% | 93-119% |