| 48T | ¥2240.00 |
| 96T | ¥3200.00 |
| 96T*5 | ¥13600.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.037 | 1.942 |
| 5.00 | 1.638 | 1.543 |
| 2.50 | 1.153 | 1.058 |
| 1.25 | 0.949 | 0.854 |
| 0.63 | 0.581 | 0.486 |
| 0.31 | 0.340 | 0.245 |
| 0.16 | 0.240 | 0.145 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-107% | 97% |
| EDTA plasma(n=5) | 89-105% | 97% |
| Heparin plasma(n=5) | 91-104% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-115% | 93-118% | 87-94% | 90-98% |
| EDTA plasma(n=5) | 85-115% | 91-113% | 88-94% | 85-92% |
| Heparin plasma(n=5) | 93-114% | 92-105% | 85-107% | 91-117% |