| 48T | ¥2240.00 |
| 96T | ¥3200.00 |
| 96T*5 | ¥13600.00 |
| Concentration (µg/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.150 | 2.056 |
| 50.00 | 1.659 | 1.565 |
| 25.00 | 1.198 | 1.104 |
| 12.50 | 0.897 | 0.803 |
| 6.25 | 0.656 | 0.562 |
| 3.13 | 0.324 | 0.230 |
| 1.56 | 0.211 | 0.117 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-116% | 102% |
| EDTA plasma(n=5) | 82-87% | 85% |
| Heparin plasma(n=5) | 82-107% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-120% | 95-106% | 84-119% | 86-103% |
| EDTA plasma(n=5) | 88-111% | 92-105% | 94-118% | 81-102% |
| Heparin plasma(n=5) | 94-107% | 87-104% | 91-106% | 82-94% |