产品中心

人脂肪酶成熟因子2(LMF2)微量上样酶联免疫吸附检测试剂盒
一键复制产品信息
ELK5126MS
规格: 价格:
48T ¥1960.00
96T ¥2800.00
96T*5 ¥11900.00

Overview

Product name: Human LMF2(Lipase Maturation Factor 2) Microsample ELISA Kit
Reactivity: Human
Alternative Names: TMEM153; TMEM112B; Transmembrane Protein 153
Assay Type: Sandwich
Sensitivity: 0.06 ng/mL
Standard: 10 ng/mL
Detection Range: 0.16-10 ng/mL
Sample Type: tissue homogenates, cell lysates and other biological fluids
Assay Length: 3.5h
Research Area: Metabolic pathway;
Uniprot ID: Q9BU23
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human LMF2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human LMF2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human LMF2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human LMF2 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (ng/mL) OD Corrected OD
10.00 2.285 2.186
5.00 1.568 1.469
2.50 1.085 0.986
1.25 0.887 0.788
0.63 0.595 0.496
0.31 0.179 0.080
0.16 0.208 0.109
0.00 0.099 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant LMF2 and the recovery rates were calculated by comparing the measured value to the expected amount of LMF2 in samples.
Matrix Recovery range Average
serum(n=5) 88-107% 98%
EDTA plasma(n=5) 83-122% 103%
Heparin plasma(n=5) 85-102% 94%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LMF2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 82-110% 88-112% 93-99% 84-101%
EDTA plasma(n=5) 93-110% 89-107% 94-115% 94-107%
Heparin plasma(n=5) 91-116% 94-102% 95-107% 84-90%

产品成分表