| 48T | ¥1960.00 | 
| 96T | ¥2800.00 | 
| 96T*5 | ¥11900.00 | 
| Concentration (μg/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.195 | 2.102 | 
| 5.00 | 1.593 | 1.500 | 
| 2.50 | 1.185 | 1.092 | 
| 1.25 | 0.852 | 0.759 | 
| 0.63 | 0.535 | 0.442 | 
| 0.32 | 0.387 | 0.294 | 
| 0.16 | 0.230 | 0.137 | 
| 0.00 | 0.093 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 94-105% | 98% | 
| EDTA plasma(n=5) | 95-107% | 101% | 
| Heparin plasma(n=5) | 80-93% | 86% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 87-102% | 85-94% | 87-101% | 89-99% | 
| EDTA plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% | 
| Heparin plasma(n=5) | 88-103% | 83-94% | 92-105% | 89-97% | 

 