| Concentration (nmol/L) | OD | Corrected OD | 
|---|---|---|
| 30.00 | 0.191 | |
| 15.00 | 0.389 | |
| 7.50 | 0.575 | |
| 3.75 | 0.895 | |
| 1.88 | 1.197 | |
| 0.94 | 1.537 | |
| 0.47 | 1.793 | |
| 0.00 | 2.211 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 79-93% | 86% | 
| EDTA plasma(n=5) | 90-105% | 97% | 
| Heparin plasma(n=5) | 86-99% | 92% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 82-95% | 83-96% | 88-95% | 96-102% | 
| EDTA plasma(n=5) | 85-100% | 84-98% | 95-101% | 88-102% | 
| Heparin plasma(n=5) | 85-94% | 90-99% | 95-102% | 82-90% | 

 