| 48T | ¥1680.00 | 
| 96T | ¥2400.00 | 
| 96T*5 | ¥10200.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1000.00 | 2.179 | 2.075 | 
| 500.00 | 1.828 | 1.724 | 
| 250.00 | 1.436 | 1.332 | 
| 125.00 | 1.036 | 0.932 | 
| 62.50 | 0.794 | 0.690 | 
| 31.25 | 0.423 | 0.319 | 
| 15.63 | 0.228 | 0.124 | 
| 0.00 | 0.104 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 85-97% | 91% | 
| EDTA plasma(n=5) | 80-94% | 87% | 
| Heparin plasma(n=5) | 91-105% | 98% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 95-102% | 93-101% | 81-89% | 85-99% | 
| EDTA plasma(n=5) | 88-97% | 79-92% | 81-97% | 87-95% | 
| Heparin plasma(n=5) | 85-94% | 96-105% | 87-101% | 82-90% | 
 