| 48T | ¥1960.00 | 
| 96T | ¥2800.00 | 
| 96T*5 | ¥11900.00 | 
| Concentration (ng/mL) | OD | Corrected OD | 
|---|---|---|
| 10.00 | 2.258 | 2.170 | 
| 5.00 | 1.593 | 1.505 | 
| 2.50 | 1.136 | 1.048 | 
| 1.25 | 0.847 | 0.759 | 
| 0.63 | 0.518 | 0.430 | 
| 0.32 | 0.385 | 0.297 | 
| 0.16 | 0.226 | 0.138 | 
| 0.00 | 0.088 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 82-95% | 88% | 
| EDTA plasma(n=5) | 87-101% | 94% | 
| Heparin plasma(n=5) | 87-99% | 93% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 83-96% | 93-102% | 95-103% | 93-102% | 
| EDTA plasma(n=5) | 82-96% | 92-103% | 90-99% | 89-102% | 
| Heparin plasma(n=5) | 85-96% | 82-96% | 85-98% | 95-104% | 
 