Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.266 | 2.169 |
500.00 | 1.647 | 1.550 |
250.00 | 1.211 | 1.114 |
125.00 | 0.952 | 0.855 |
62.50 | 0.567 | 0.470 |
31.25 | 0.389 | 0.292 |
15.63 | 0.226 | 0.129 |
0.00 | 0.097 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 92-106% | 99% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 78-92% | 85% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-96% | 81-97% | 84-97% | 88-97% |
EDTA plasma(n=5) | 85-92% | 79-96% | 87-103% | 88-103% |
Heparin plasma(n=5) | 81-98% | 78-90% | 95-102% | 95-103% |