Product name: DHA(Docosahexaenoic Acid) ELISA Kit
Reactivity: General
Alternative Names: Doconexent; Cervonic Acid
Assay Type: Competitive Inhibition
Sensitivity: 4.79 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: Serum, plasma and other biological fluids
Assay Length: 2h
Research Area: Enzyme & Kinase;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with DHA. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to DHA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of DHA in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
1000.00 0.183
500.00 0.310
250.00 0.567
125.00 0.842
62.50 1.139
31.25 1.537
15.63 1.869
0.00 2.157

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant DHA and the recovery rates were calculated by comparing the measured value to the expected amount of DHA in samples.
Matrix Recovery range Average
serum(n=5) 82-95% 88%
EDTA plasma(n=5) 80-93% 86%
Heparin plasma(n=5) 95-108% 90%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of DHA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-97% 91-105% 95-107% 82-96%
EDTA plasma(n=5) 81-94% 93-107% 93-102% 91-105%
Heparin plasma(n=5) 92-104% 91-101% 87-98% 89-103%