ELK7899
规格: 价格:
48T ¥2240.00
96T ¥3200.00

Overview

Product name: SCT(Salmon Calcitonin) ELISA Kit
Reactivity: General
Alternative Names: CGRP; Salcatonin
Assay Type: Competitive Inhibition
Sensitivity: 31 pg/mL
Standard: 20000 pg/mL
Detection Range: 312.5-20000 pg/mL
Sample type: biological agents
Assay length: 2h
Research Area: Metabolic pathway;Endocrinology;Hormone metabolism;Bone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with SCT. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to SCT. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of SCT in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
20000.00 0.197
10000.00 0.332
5000.00 0.596
2500.00 0.842
1250.00 1.168
625.00 1.537
312.50 1.786
0.00 2.028

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant SCT and the recovery rates were calculated by comparing the measured value to the expected amount of SCT in samples.
Matrix Recovery range Average
serum(n=5) 93-107% 100%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 87-99% 88%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of SCT and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-105% 82-95% 81-96% 83-94%
EDTA plasma(n=5) 80-92% 86-102% 87-103% 85-101%
Heparin plasma(n=5) 85-92% 87-101% 89-99% 85-95%