Product name: | Mouse NT-ProBNP(N-Terminal Pro-Brain Natriuretic Peptide) ELISA Kit |
Reactivity: | Mouse |
Alternative Names: | NT-Pro-BNP; ; N-BNP; ANFB_HUMAN; BNP(1-32); BNP(5-29); BNP-32; Gamma-brain natriuretic peptide; nppb |
Assay Type: | Sandwich |
Sensitivity: | 28 pg/mL |
Standard: | 5000 pg/mL |
Detection Range: | 78.13-5000 pg/mL |
Sample type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Assay length: | 3.5h |
Research Area: | Endocrinology;Cardiovascular biology; |
Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse NT-ProBNP. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse NT-ProBNP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse NT-ProBNP, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse NT-ProBNP in the samples is then determined by comparing the OD of the samples to the standard curve. |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
5000.00 | 2.103 | 2.026 |
2500.00 | 1.608 | 1.531 |
1250.00 | 1.233 | 1.156 |
625.00 | 0.937 | 0.860 |
312.50 | 0.527 | 0.450 |
156.25 | 0.361 | 0.284 |
78.13 | 0.206 | 0.129 |
0.00 | 0.077 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 86-99% | 92% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 90-105% | 97% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-100% | 84-98% | 95-101% | 88-102% |
EDTA plasma(n=5) | 88-104% | 84-98% | 95-105% | 88-102% |
Heparin plasma(n=5) | 87-96% | 91-101% | 87-98% | 89-103% |