产品中心

人胰高血糖素样肽2(GLP2)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK3108
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Human GLP2(Glucagon Like Peptide 2) ELISA Kit
Reactivity: Human
Alternative Names: Glp2; GLP2
Assay Type: Competitive Inhibition
Sensitivity: 50.2 pg/mL
Standard: 10000 pg/mL
Detection Range: 156.25-10000 pg/mL
Sample type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length: 2h
Research Area: Endocrinology;Cardiovascular biology;Hepatology;Hormone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human GLP2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human GLP2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human GLP2 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
10000.00 0.199
5000.00 0.357
2500.00 0.575
1250.00 0.823
625.00 1.169
312.50 1.585
156.25 1.796
0.00 2.112

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant GLP2 and the recovery rates were calculated by comparing the measured value to the expected amount of GLP2 in samples.
Matrix Recovery range Average
serum(n=5) 85-99% 92%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 93-107% 100%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GLP2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-104% 82-96% 92-101% 82-98%
EDTA plasma(n=5) 85-99% 98-106% 86-99% 78-95%
Heparin plasma(n=5) 85-102% 89-103% 97-105% 91-99%
关闭