产品中心

人凝血因子Ⅱ(F2)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK1634
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Human F2(Coagulation Factor II) ELISA Kit
Reactivity: Human
Alternative Names: FII; TM; PT; Thrombin; Prothrombin; Pro-Thrombin; Activation peptide fragment 1; Activation peptide fragment 2
Assay Type: Sandwich
Sensitivity: 0.19 ng/mL
Standard: 300 ng/mL
Detection Range: 4.69-300 ng/mL
Sample type: Serum
Assay length: 3.5h
Research Area: Signal transduction;Hematology;
Uniprot ID: P00734
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human F2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human F2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human F2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human F2 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (ng/mL) OD Corrected OD
300.00 2.132 2.049
150.00 1.593 1.510
75.00 0.935 0.852
37.50 0.752 0.669
18.75 0.543 0.460
9.38 0.357 0.274
4.69 0.257 0.174
0.00 0.083 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant F2 and the recovery rates were calculated by comparing the measured value to the expected amount of F2 in samples.
Matrix Recovery range Average
serum(n=5) 80-95% 87%
EDTA plasma(n=5) 86-99% 92%
Heparin plasma(n=5) 80-95% 87%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-104% 82-96% 92-105% 83-92%
EDTA plasma(n=5) 87-93% 85-95% 89-99% 85-95%
Heparin plasma(n=5) 89-97% 85-94% 79-96% 85-99%