产品中心

大鼠β-骨胶原交联(bCTx)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK1546
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Rat bCTx(Beta-Crosslaps) ELISA Kit
Reactivity: Rat
Alternative Names: Beta CTx; b-CTx; bCTXI; bCTX-I; B-Cr; BCL; Type I Collagen C-Telopeptide-Related Fraction
Assay Type: Competitive Inhibition
Sensitivity: 39.1 pg/mL
Standard: 8000 pg/mL
Detection Range: 125-8000 pg/mL
Sample type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length: 2h
Research Area: Bone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Rat bCTx. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat bCTx. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat bCTx in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
8000.00 0.175
4000.00 0.367
2000.00 0.583
1000.00 0.857
500.00 1.156
250.00 1.511
125.00 1.799
0.00 2.115

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant bCTx and the recovery rates were calculated by comparing the measured value to the expected amount of bCTx in samples.
Matrix Recovery range Average
serum(n=5) 82-96% 89%
EDTA plasma(n=5) 78-90% 84%
Heparin plasma(n=5) 85-97% 91%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of bCTx and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-99% 82-96% 92-101% 82-90%
EDTA plasma(n=5) 79-92% 80-93% 81-98% 89-102%
Heparin plasma(n=5) 87-98% 92-104% 82-96% 95-102%
关闭

在线咨询

Online consultation

  • 在线咨询
  • 技术支持

关注微信公众号

微信扫一扫立即咨询

微信扫一扫立即咨询

关闭