产品中心

小鼠β-骨胶原交联(bCTx)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK1545
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Mouse bCTx(Beta-Crosslaps) ELISA Kit
Reactivity: Mouse
Alternative Names: Beta CTx; b-CTx; bCTXI; bCTX-I; B-Cr; BCL; Type I Collagen C-Telopeptide-Related Fraction
Assay Type: Competitive Inhibition
Sensitivity: 27.5 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample type: Serum, plasma, urine and other biological fluids
Assay length: 2h
Research Area: Bone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Mouse bCTx. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse bCTx. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse bCTx in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
5000.00 0.191
2500.00 0.396
1250.00 0.611
625.00 0.833
312.50 1.197
156.25 1.583
78.13 1.762
0.00 2.123

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant bCTx and the recovery rates were calculated by comparing the measured value to the expected amount of bCTx in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 80-95% 87%
Heparin plasma(n=5) 90-107% 98%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of bCTx and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 92-104% 93-99% 87-953% 82-93%
EDTA plasma(n=5) 88-96% 90-101% 93-102% 85-94%
Heparin plasma(n=5) 79-91% 89-103% 82-95% 86-99%