产品中心

三碘甲状腺原氨酸(T3)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK1339
规格: 价格:
48T ¥1680.00
96T ¥2400.00

Overview

Product name: T3(Triiodothyronine) ELISA Kit
Reactivity: General
Alternative Names: 3,3',5-Triiodo-L-Thyronine; Liothyronine; Cytomel; Tertroxin
Assay Type: Competitive Inhibition
Sensitivity: 52.7 pg/mL
Standard: 10000 pg/mL
Detection Range: 156.25-10000 pg/mL
Sample type: serum, plasma and other biological fluids
Assay length: 2h
Research Area: Endocrinology;Autoimmunity;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with T3. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to T3. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of T3 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
10000.00 0.197
5000.00 0.385
2500.00 0.575
1250.00 0.812
625.00 1.197
312.50 1.592
156.25 1.793
0.00 2.226

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant T3 and the recovery rates were calculated by comparing the measured value to the expected amount of T3 in samples.
Matrix Recovery range Average
serum(n=5) 87-99% 93%
EDTA plasma(n=5) 82-98% 90%
Heparin plasma(n=5) 90-107% 98%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of T3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 95-104% 82-96% 92-103% 90-99%
EDTA plasma(n=5) 91-102% 86-102% 87-99% 99-105%
Heparin plasma(n=5) 86-93% 91-101% 87-98% 95-104%